IMED/Gemini/Alaris Pump & Tubing

Shopping Cart
0 items
Additional Links
DOTmed Certified: HR Medical, Inc.
IMED/Gemini/Alaris Pump & Tubing
IMED/Gemini/Alaris Pump & Tubing

 

Alaris/Cardinal Imed Gemini PC-1 Infusion Pump

For Sale or Rent 

Product Specifications

Operating Principle:
Linear peristalsis

 

Dimensions:
10.8" x 8.2" x 6.5" (27.5 cm x 20.6 cm x 16.5 cm)

 

Weight:
11.3 lbs (5.1 kg)

 

Power Requirements:
90 - 130 (120 nominal) VA, 0.5A, 50/60 Hz, 3 wire, single phase

 

Battery Operation:
With a new battery, approximately 5 hours At 125 ml/hr before a "HELP BATTERY"

 

Electrical Leakage:
Less than 100 microamps

 

Delivery Rate Range

Micro:
0.1-999 ml/hr (Note: Recommended maximum rate in the Controller Mode is 500 ml/hr

 

Macro:
0.1-99.9 ml/hr (in 0.1 ml/hr increments)

 

Volume to be Infused

Micro:
0.1-999.9 m/l (in 0.1 ml increments)

 

Macro:
1-9999 ml (in 1 ml increments)

 

KVO (Keep Vein Open) Rate:
1 ml/hr if set rate is 1 ml/hr or above, or actual rate if set at 0.9 ml/hr or below

 

Occlusion Pressure

Pump mode:
10 ±2 psi for rates above 30 ml/hr. For rates of 30 ml/hr and below, the occlusion pressure will be less than 10 psi, and is rate dependent (unless locked into 10 psi for all rates)

 

Advisories:
Secondary, Primary, Infusion Complete - KVO, Empty Container-KVO, Low Battery, Maintenance, Pause, On Batt, 10 psi. Flow, Total Vol Infused, Sec Vol Infused Macro, Micro, Versataper: Setup, Step: xx, Steps: xx, Total Vol: xx, Duration: xx, Versataper, and Time Left:xx

 

Prompts:
Select P/C mode, set rate, set VTBI, and Press Start

 

Alarms:
Check ECD, Check IV Set, Close Door, Air-in-Line, Occluded Patient Side, Occluded Fluid Side, Occluded, Partial Occlusion Fluid Side, Flo-Stop Open/Close Door, KVO Comm Cable, KVO Comm Tout, and KVO

 

Malfunctions:
Help internal error, help battery, and (Watchdog Malfunction)

 

Secondary (Piggyback):
Dual rate programming 